Calibration Of Glassware

Topics: Experiment

This experiment must be repeated three times or more and take the average reading to get more accurate data. By using the formulae, the standard deviation and relative standard deviation could be calculated. As the result, the lower the standard deviation, the higher the consistency of an instrument. Although the instrument may measure a given sample, the value might need to be altered due to various conditions and errors. Objectives The objective of the experiments was to; Calibrate a 10 ml volumetric pipette.

Calibrate a 25 ml volumetric pipette. Calibrate a 100 ml volumetric flask.

Calibrate a 50 ml measuring cylinder Introduction Volumetric glassware is used to accurately measure volumes. Despite all of the tools and instruments available for the use of measuring values of various data, it is impossible to measure the true value of anything. This experiment is intended to develop a person’s capacity to handle volumetric glassware. One must understand how to handle volumetric glassware to acquire the best possible data from the equipment used in lab.

Although the instrument may measure a given sample, the value might need to be altered due to various conditions.

However, with practice and experience, we can reach very close to the true value, gaining the ability to measure with both high accuracy and precision. The objective of this experiment is to calibrate a ml burette and other volumetric glassware by calculating their correction values. By doing so, we can correct systematic errors caused by the burette’s or glassware’s values. A systematic error is an error in reading the scale when a physical quantity is being measured.

Get quality help now
KarrieWrites
Verified

Proficient in: Experiment

5 (339)

“ KarrieWrites did such a phenomenal job on this assignment! He completed it prior to its deadline and was thorough and informative. ”

+84 relevant experts are online
Hire writer

Systematic errors are caused by the instrument, observer and surroundings.

Calibration Of Volumetric Pipette

Another main type of error is random error. Random error is an error which occurs when the observer is reading the scale on the measuring instrument. Volumetric glassware is used to accurately measure volumes. In any given experiment, students may come across systematic errors or random errors. Random errors are not able to be corrected by a standard mean because they are not consistent. However, systematic errors are consistent and is able to be detected which allows students to correct the error that was found. So, using uncelebrated glassware will cause a systematic error.

This error occurs because he volume reading given by the glassware is slightly different than the actual volume that is obtained in the glassware. In order to avoid systematic error, it important to calibrate the measuring instrument so that there is a standard value for zero which allows the measurement to be more accurate. Calibration of volumetric glassware such as burette is carried out by weighing the amount of water delivered. The temperature of the water used in the calibration must be known since the density of water changes.

The purpose of this experiment is to investigate the measurement of the actual illume contents of volumetric glassware. If there is a slight variation in this instrument, it would cause systematic error and thus, the accuracy of results of an experiment or investigation would be affected. All volumetric glassware has marking on it. It is either being marked with AD or ETC, which brings the meaning of ‘to deliver’ and ‘to contain’ respectively. For example, a pipette has a marking of AD, which means a pipette is marked as a ‘to deliver’.

Whereas, volumetric flask has ETC as its marking, meaning it is marked as a ‘to contain’. A volumetric pipette measuring cylinder can be calibrated by just weighing the water they deliver. As for volumetric flask, the weight of an empty flask is recorded. Next, weigh the flask after filling it with water to the mark. After that, it is crucial to convert the mass to volume. The water density at a temperature will aid this process. The compliance of the Volume Occupied by 1. Egg of Water Weigh in Air table is deemed necessary throughout the comparison.

In order to obtain the true volume of volumetric glassware holds, this formula will be used. This experiment shall be repeated twice or more to enhance accuracy of the exults obtained. Besides detecting systematic errors, this experiment would aid on the technique and understandings to the correct use of these equipments. Materials Transfer pipette, cleaning solution/detergent, beaker, thermometer, distilled water, top loading balance, ml volumetric pipette, millimeters pipette, ml volumetric flask, ml measuring cylinder.